Welcome to Fonar.com
 
Home » Products & Services » Upright MRI » Siting
Upright (TM) MRI Specifications
Siting Requirements

Magnet Room

Minimum Installed Area 504 sq. feet [28' x 18'] (46.8 sq. meters)
Ceiling Height Required 11 feet (3.35 meters)
Magnet Specification

Weight

145.5 tons (291,000 lbs.) Assembled in pieces on-site

Frame Outer Dimensions

102 in. (259cm) W x 107 in.(272 cm) H x 176 in. (447 cm) D

Power Requirements

480 Volt, 3-phase, 125 Amps

Power Consumption

100 KVA, The magnet is turned off when not in use
Magnet Room Heat Load: 30,000 BTU

Equipment Room

Minimum Installed Area 240 sq. feet [12' x 20'] (21 sq. meters)
Heat Load 48,000 BTU
System Racks

Power Requirements:

120/208 Volt, 3-phase, 200 Amps

Power Consumption:

20 KVA
Closed-loop air-cooled water chiller (included with system)

Capacity

30 tons

Power Requirements

480 Volt, 3-phase, 100 Amps

Power Consumption

47 KVA (when running)
top of page